OPT104-Suspended Solids Checker (TCS-1000)- Made in Japan

Category: Water

 
 
This checker can be used for continuous monitoring of the water turbidity (TC-100/500/3000) and suspended solids (TCS-1000) .
  Analogue output (4-20mA) allows you to use this checker with various equipments.
Wiper for easy maintenance
  Original wiper certainly keeps lenses clean.
Stain of lenses need to be checked only once a month.
Compact size
  Simple cylindrical design in compact size approx ¢32 x 170mm (1.26 x 6.69 inch).
Requires small space to set up the unit.
High durability
  Durable stainless steel has been used for the sensor casing.
Stainless steel = 18-15-2.5 Cr-Ni-Mo low C stainless steel
Lenses are made of hard-to-scratch sapphire glass.

TC-100/TC-500/TC-3000
Turbidity Checker
TC-100/500/3000
SS Checker
TCS-1000
The transmitter is also available to use in combination with the Turbidity Checker / SS Checker.
  As well as the turbidity/SS indication, this transmitter can indicate signal output response time, alarm level setting and so on.
You can control the turbidity/SS utilizing its functions fully.

Power Supply
50/60Hz
Signal output
Alarm output
Degree of protection Jet-proof type (IP65)

SC-T3
Turbidity Checker Transmitter
SC-T4
SS Checker Transmitter
Model No. TC-100 TC-500 TC-3000
Measuring range 0-100 (FTU) 0-500 (FTU) 0-3000 (FTU)
Power supply DC12V±10%
Current Draw At regular operation: 30mA (Max.), At cleaning operation: 240mA (Max.) (excluding analog signal output)
Output Analog (4-20mA) signal output: resistance load of 300(Max.)
4mA: 0 FTU
20mA: 100 FTU (TC-100), 500 FTU (TC-500), 3000 FTU (TC-3000)
Self-checking output: open collector (12VDC 20mA Max.)
Input Calibration signal input
Cleaning system Both sided swing wiper cleaning system
Time interval for cleaning Clean once immediately after power-on, andsubsequently clean once every 30 minutes Clean once immediately after power-on, and subsequently clean once every 10 minutes
Operating temperature 0-40°C (unfrozen)
Major material SUS 316L, Sapphire glass, Fluorocarbon rubber, EPDM, POM
Dimensions Ø 32x163mm (excluding horizontal hanging attachment)
Weight approx. 930g
Degree of Protection IP 68, maximam depth of 2 meters (underwater type)
Detector cable length 10m
Option Transmitter (SC-T3), Mounting attachment (TA-1), Maintenance kit (TC-MK)
Model No. TCS-1000
Measuring range 0-1000 (mg/l) (standard substance:kaolin)
Power supply DC12V±10%
Current draw At regular operation: 30mA (Max.), At cleaning operation: 240mA (Max.)
(excluding analog signal output)
Output Analog (4-20mA) signal output: resistance load of 300Ω(Max.)
4mA: 0 (mg/l)
20mA: 1000 (mg/l)
Self-checking output: open collector (12VDC 20mA Max.)
Input calibration signal input
Cleaning system Both sided swing wiper cleaning system
Time interval for cleaning Clean once immediately after power-on, and subsequently clean once every 10 minutes.
Operating temperature 0-40°C (unfrozen)
Major material SUS 316L, sapphire glass, fluorocarbon rubber, EPDM, POM
Dimensions Ø 32x170mm (excluding horizontal hanging attachment)
Weight approx. 930g
Degree of Protection IP 68, maximum depth of 2 meters (underwater type)
Detector cable length 10m
Option Transmitter (SC-T4), Mounting attachment (TA-1), Maintenance kit (TC-MK)
Note: Specifications and design are subject to change without prior notice.
Model No. SC-T3(For TC-100/500/3000)
Power Supply AC100 - 240V±10% 50/60Hz
Voltage Draw At regular operation: 8VA (Max.)
At cleaning operation: 16VA (Max.)
(including analog signal input/output 20mA)
Display Resolution TC-100: 0.1 TC-500: 1 TC-3000: 5
Output Calibration output
Power supply for detector (12 VDC)
Analog (4-20mA) signal output: resistance load of 300 Max.
Self-checking output (non-voltage C-contact capacity 240 VAC, 1A resistance load)
Alarm output (non-voltage C-contact capacity 240VAC 1A resistance load)
Input (detector) Detector signal input (analog 4-20mA, input resistance approx.100)Self-checking input
Operating Temperature -20~+50°C Humidity: 95%Rh Max. (Not exposed to direct sunlight)
Major Material Polycarbonate
Dimensions HxWxD= approx. 162X240X75 mm
Degree of Protection IP65 (jetproof type)

Option

Hood kit (TP-FK2), Pole stanchion (PS-1)

Model No. SC-T4(For TCS-1000)
Power 50/60Hz
Power consumption Normal: 8VA or less, during washing: 16VA or less
(including detector, at the time of analog signal input/output of 20mA)
Display resolution 0.1mg/l(0-49.9mg/1)
1mg/l(50-1000mg/l)
Output Calibration Output
Power Supply for Sensor (12 VDC)
Signal Output (analog 4-20mA, resistance load of 300Ω or less)
Self-checking Relay Output (non-voltage C-contact capacity 240 VAC, 1A resistance load)
Alarm Relay Output (non-voltage C-contact capacity 240 VAC, 1A resistance load)
Input (detector) Detector signal input (analog 4-20mA, input resistance approx.100)
Self-checking input
Operating temperature range -20 to +50 C・, humidity 95%Rh or less (Avoid direct sunlight)
Main material Polycarbonate
Size HxWxD = approx.162x240x75 mm
Mass Approx. 1.6kg (including Mounting Brackets)
Protective structure IP65 (jetproof type)

 
TC-100/500/3000 TCS-1000
  TC-100/500/3000 TCS-1000
  SC-T3 / SC-T4
  SC-T3

Location

Ask about this product